Home / Products / GS850 cold-seal autoclave
GS850 AUTOCLAVE / COLD-SEAL AUTOCLAVE

GS850 Cold-Seal Autoclave Ore-Formation Simulation System

A batch-experiment platform for ore-formation simulation and high-pressure high-temperature water-rock reaction. 300MPa maximum working pressure, 850℃ maximum operating temperature, with magnet-ring quench, pressure-hold leak testing and programmed pressure-temperature curves under full-process control. Single-unit customisation and multi-unit parallel operation are supported.

01

Overview

The GS850 cold-seal autoclave ore-formation simulation system is developed and manufactured by Hubei Guangshi Precision Co., Ltd. It reproduces the temperature and pressure conditions of low-to-medium temperature hydrothermal environments in the laboratory, for ore-formation process simulation, water-rock reaction and long-duration constant-temperature experiments. The system uses a cold-seal vessel structure: the vessel carries the pressure while the furnace controls temperature independently. The cap is tightened with a torque wrench to a set value (≥80 N·m), and the double-cone sealing faces deliver 300MPa-class closure.

The system is equipped with a magnet-ring quench mechanism: at the end of an experiment, pulling the magnet ring moves the sample rack to the cold zone in one motion. A pressure rise of 40–50MPa confirms a successful quench, so samples can be quenched rapidly without breaking the seal, preserving intermediate reaction states.

Max working pressure
300MPa
Max operating temperature
850℃
Pressure-hold leak test
30–60 min · ≤2MPa
Quench criterion
Pressure rise 40–50MPa
02

Technical specifications

ItemSpecificationNotes
System nameGS850 cold-seal autoclave ore-formation simulation systemUnit numbering SB-SR-001~010
Maximum working pressure300MPaAs marked on the equipment nameplate
Maximum operating temperature850℃As marked on the equipment nameplate
Sealing structureCold-seal · double-cone sealCap torque ≥80 N·m; hose 16–20 N·m (18 N·m recommended)
Pressure-hold leak test30–60 min · fluctuation ≤2MPaLess than 30 min is not enough to expose micro-leaks
Pre-heating depressurisation≥1/4 of the experimental pressurePrevents overpressure from thermal expansion on heating; alternatively depressurise fully and re-pressurise
Daily inspection criterionPressure drop ≤1%/24h is normalA drop ≥1MPa within 1h indicates an abnormal leak
Leak detectionWith fluid: bake at 120℃ for 2h and weigh
Without fluid: immerse in alcohol for 2h and weigh
Any weight change means a leak; do not put into service
Quench methodMagnet-ring quenchA pressure rise of 40–50MPa confirms success; no rise means failure
Depressurisation temperature limit<100℃Above 100℃, opening valves to depressurise or disassembling is strictly prohibited
Cooling methodForced cooling by circulating water chillerMust run continuously throughout the entire heating phase
Run controlProgrammed pressure-temperature curvesMulti-segment heating and soaking programmes, logged run by run
These values match the equipment nameplate and the GS850 Equipment Experiment Operating Procedure, and are the single site-wide standard.
03

Safety & operation

The GS850 is governed by both the GS850 Equipment Experiment Operating Procedure and the quick-start manual. The core discipline is check water first, check the station second, never heat without a passed pressure hold, never touch a wrench above 100℃:

  • Five-step workflow: pre-checks → sample loading and sealing → pressurisation and pressure hold → heating and soak → quench, cool and sample retrieval
  • Two-person red line: sample loading, sampling and any work under pressure must never be done by a single operator; two people must be present
  • Batch supervision: one operator may supervise at most 10 vessels during daily inspection; all readings must be taken at the same moment, from the same gauge
  • No work under pressure: never tighten or loosen any connection (hose, cap, valve) while pressurised
  • Interlock conditions: never start heating without the circulating chiller running; never enter the heating phase without releasing ≥1/4 of the pressure first
  • Over-limit protection: never bypass or short-circuit overpressure/overtemperature alarms or safety valves; never operate beyond nameplate ratings
  • Full record-keeping: checks, pressure holds, inspections and quench results are logged entry by entry, with one run record per vessel archived with the experiment

Abnormal-leak response

If pressure-hold fluctuation exceeds 2MPa or pressure drops ≥1MPa within 1h: depressurise to atmospheric → inspect the cone sealing faces and fittings → re-pressurise and re-test; if pressure cannot be maintained, follow the early-termination procedure to cool and retrieve samples. Never tighten fittings under pressure.
04

Applicable experiments

01

Ore-formation process simulation

Simulate the migration, metasomatism and mineral precipitation of ore-bearing hydrothermal fluids through host rock, and study the mechanisms of element transport, enrichment and spatial zonation.

02

Water-rock reaction kinetics

Study rock alteration, dissolution–precipitation equilibrium and element transport rates under long-duration high-pressure high-temperature conditions.

03

Hydrogen generation simulation

Study hydrogen generation from supercritical water–rock interaction, supported by the methodology of our hydrogen-generation simulation white paper.

04

Material corrosion assessment

Evaluate the corrosion behaviour and service life of alloys, coatings and sealing materials in high-pressure high-temperature aqueous environments.

05

Purchase & contract services

The GS850 is available in two cooperation modes: complete-unit purchase and contract experiment services. Contract experiments are billed at ¥50 per machine-hour; pressure-temperature programmes are tailored to your project, and experiment records and samples are delivered with the order:

STEP 01
Conditions review
STEP 02
Proposal & quotation
STEP 03
Scheduling
STEP 04
Delivery & archiving
  • Complete-unit delivery includes the technical manual, quick-start manual and on-site training
  • Pressure and temperature are calibrated by actual measurement before delivery, with the figures shipped with the unit
  • Contract experiments follow a unified machine-hour basis; see contract experiment services
  • National service hotline 400-184-9193 · humin@gsjgmetal.cn

Tell us your experimental conditions

Target temperature and pressure range, fluid composition, sample type and loading, soak duration and quench requirements — the more specific the information, the better we can tailor the configuration to your needs.
06

FAQ

How do I know the magnet-ring quench succeeded?

Pull the magnet ring to move the sample rack to the cold zone in one motion. A pressure rise of 40–50MPa confirms a successful quench; no rise means the quench failed. The seal stays intact throughout, so intermediate reaction states are preserved.

Why must the pressure hold last 30–60 minutes?

Micro-leaks take time to show up. A hold shorter than 30 minutes is not sufficient to expose them. The hold passes only when fluctuation stays within 2MPa over 30–60 minutes; otherwise the vessel must be depressurised, the cone sealing faces and fittings inspected, and the test repeated.

How is a leak detected during daily operation?

During routine inspection, a pressure drop of ≤1%/24h is normal; a drop ≥1MPa within 1h indicates an abnormal leak. For a definitive check: with fluid inside, bake at 120℃ for 2h and weigh; without fluid, immerse in alcohol for 2h and weigh — any weight change means a leak and the vessel must not be used.