Home / Products / GS1100 cold-seal autoclave
GS1100 AUTOCLAVE / MOLYBDENUM-BASED VESSEL · 1100℃

GS1100 Cold-Seal Autoclave Into the 1100℃ High-Temperature Range

A molybdenum-based vessel platform for high-temperature ore-formation simulation and melt–fluid reaction. Maximum operating temperature 1100℃, working pressure 300MPa, with argon protective atmosphere, furnace-chamber evacuation, two-stage pressurisation and magnet-ring quench under full-process control — extending the upper temperature limit of the GS series.

01

Overview

The GS1100 cold-seal autoclave ore-formation simulation system is developed and manufactured by Hubei Guangshi Precision Co., Ltd. It uses a molybdenum-based vessel to extend the upper temperature limit of the GS series to 1100℃, for high-temperature ore-formation simulation, melt–fluid interaction and deep geochemistry experiments.

Because of the metallurgical character of molybdenum-based alloys, the system carries an independent set of high-temperature safeguards: the vessel may only be pressurised once ambient temperature is above 20℃ and the vessel has been preheated to 100℃ (pressurising a cold vessel causes brittle fracture); during heating, the furnace chamber is evacuated first and then swept with argon at 0.04–0.05MPa, low-rate and continuous, to protect the vessel and the sample; pressurisation is two-stage — a first stage at 20–40MPa, then a second-stage pump up to the pressure specified by the project.

Vessel material
Molybdenum-based alloy
Max operating temperature
1100℃
Working pressure
300MPa
Protective atmosphere
Argon 0.04–0.05MPa
02

Technical specifications

ItemSpecificationNotes
System nameGS1100 cold-seal autoclave ore-formation simulation systemUnit numbering SB-SR-011~020
Vessel materialMolybdenum-based alloyAmbient temperature must be above 20℃; pressurising cold is strictly prohibited
Maximum operating temperature1100℃As marked on the equipment nameplate
Working pressure300MPaAs marked on the equipment nameplate
Pressurisation methodStage 1 at 20–40MPa + stage 2 to the specified pressureValve sequence must never be reversed
Sealing torqueCap ≥80 N·m · hose ≥15 N·mSet values with a torque wrench and record the measured values
Pressure-hold leak test30–60 min · pressure drop <2MPaIf out of tolerance, depressurise and investigate; never tighten under pressure
Pre-heating depressurisationRelease about 1/4 of the experimental pressureAfter heating starts, never vent the vessel pressure completely
Furnace atmosphereEvacuation + argon 0.04–0.05MPaEvacuation lasts only a few seconds; argon flows at a low rate throughout
Heating rateApprox. 3℃/minUse this to compute heating time; if a quench is required, extend the soak by at least one extra day
Quench methodMagnet-ring quenchA pressure rise of 40–50MPa confirms success; the air compressor must be running
Depressurisation temperature limit<100℃Above 100℃, opening valves to depressurise or disassembling is strictly prohibited
Sample fixingOnly non-magnetic molybdenum or nickel wireMagnetic wire will scrap the vessel (red line)
These values match the equipment nameplate and the GS1100 Equipment Operating Procedure, and are the single site-wide standard.
03

Safety & operation

The GS1100 is governed by both the GS1100 Equipment Operating Procedure and the quick-start manual. Because of the low-temperature brittleness of the molybdenum-based vessel and its atmosphere requirements, its operating discipline is stricter than that of ambient-temperature vessels:

  • Preheating red line: ambient temperature must be above 20℃ and the vessel preheated to 100℃ before pressurisation, to prevent low-temperature brittle fracture of the molybdenum-based vessel
  • Wire red line: fix samples only with non-magnetic metal wire such as molybdenum or nickel wire; magnetic wire will scrap the vessel
  • Pressure-hold red line: after heating starts, the vessel pressure must never be vented completely — doing so shatters the sample
  • Shutdown red line: before sampling, manually issue the stop plan to terminate the programme; never rely entirely on automatic program stop
  • Quench condition: the quench lift function is unavailable unless the air compressor is running
  • Two-person operation: two people must be present and wearing protective equipment for sample loading, sampling and any work under pressure; one operator may supervise at most 10 vessels during daily inspection
  • Acceptance gate: sample installation may begin only after every pre-check item has passed

New-vessel run-in requirement

A new autoclave must never run at limit conditions early in its life; doing so easily damages the vessel. For the first batch of experiments, keep a margin within the rated temperature and pressure range and approach the target conditions step by step.
04

Applicable experiments

01

High-temperature ore-formation simulation

Reconstruct magmatic-hydrothermal transition environments in the 1100℃ range and study mineral dissolution, transport and precipitation mechanisms at high temperature.

02

Melt–fluid reaction

Study the interaction of silicate melts with high-pressure high-temperature fluids, element partitioning and volatile behaviour.

03

Deep geochemistry

Simulate geochemical reaction paths at crustal-depth temperature and pressure conditions, supporting deep-earth process and material cycling research.

04

High-temperature material assessment

Evaluate phase transformation, corrosion and service behaviour of refractory metals, ceramics and coating materials in high-pressure high-temperature environments.

05

Purchase & contract services

The GS1100 is available in two cooperation modes: complete-unit purchase and contract experiment services. Contract experiments are billed at ¥50 per machine-hour; pressure-temperature programmes are tailored to your project, and experiment records and samples are delivered with the order:

STEP 01
Conditions review
STEP 02
Proposal & quotation
STEP 03
Scheduling
STEP 04
Delivery & archiving
  • Complete-unit delivery includes the technical manual, quick-start manual and on-site training
  • Pressure and temperature are calibrated by actual measurement before delivery, with the figures shipped with the unit
  • Contract experiments follow a unified machine-hour basis; see contract experiment services
  • National service hotline 400-184-9193 · humin@gsjgmetal.cn

Tell us your experimental conditions

Target temperature and pressure range, melt/fluid system, sample type and loading, soak duration and quench requirements — the more specific the information, the better we can tailor the configuration to your needs.
06

FAQ

Why must the GS1100 never be pressurised cold?

Molybdenum-based alloy is brittle at low temperature, so pressurising a cold vessel causes brittle fracture. Ambient temperature must be above 20℃ and the vessel preheated to 100℃ before pressurisation begins.

What wire can be used to fix samples?

Only non-magnetic metal wire — molybdenum wire or nickel wire. Magnetic wire will scrap the vessel; this is a hard red line of GS1100 operation.

Why does the furnace chamber need vacuum and argon?

At 1100℃ the molybdenum-based vessel and the sample need protection from oxidation. The chamber is evacuated first (only a few seconds of pumping), then argon flows at 0.04–0.05MPa at a low continuous rate throughout the run.